dc.date.issued2018-01-08
dc.identifier.citationKumaran , B & Watson , T 2018 , ' Skin thermophysiological effects of 448 kHz capacitive resistive monopolar radiofrequency in healthy adults : A randomised crossover study and comparison with Pulsed Shortwave Therapy ' , Electromagnetic Biology and Medicine , vol. 37 , no. 1 , pp. 1-12 . https://doi.org/10.1080/15368378.2017.1422260
dc.identifier.issn1536-8378
dc.identifier.otherORCID: /0000-0003-3125-0902/work/83087686
dc.identifier.otherORCID: /0000-0002-1332-9337/work/125979209
dc.identifier.urihttp://hdl.handle.net/2299/20411
dc.descriptionThis is an Accepted Manuscript of an article published by Taylor & Francis Group in Electromagnetic Biology and Medicine on 8 January 2018, available online: https://doi.org/10.1080/15368378.2017.1422260. Under embargo until 8 January 2019.
dc.description.abstractRadiofrequency-based electrophysical agents (EPA) have been used in therapy practice over several decades (e.g., shortwave therapies). Currently, there is insufficient evidence supporting such devices operating below shortwave frequencies. This laboratory-based study investigated the skin physiological effects of 448 kHz capacitive resistive monopolar radiofrequency (CRMRF) and compared them to pulsed shortwave therapy (PSWT). In a randomised crossover study, seventeen healthy volunteers received four treatment conditions–High, Low and Placebo dose conditions receiving 15-min CRMRF treatment and a Control condition receiving no intervention. Fifteen participants also received high dose PSWT for comparison. Treatment was applied to the right lower medial thigh. Pre, post and 20-min follow-up measurements of skin temperature (SKT), skin blood flow (SBF) and nerve conduction velocity (NCV) were obtained using Biopac MP150 system. Group data were compared using the ANOVA model. Statistical significance was set at p ≤ 0.05 (0.8P, 95%CI). Significant increase and sustenance of SKT with both high and low dose CRMRF was demonstrated over the other groups (p < 0.001). PSWT increased SKT significantly (p < 0.001) but failed to sustain it over the follow-up. However, among the five conditions, only high dose CRMRF significantly increased and sustained SBF (p < 0.001). Overall, the CRMRF physiological responses were significantly more pronounced than that of PSWT. No significant changes in NCV were noted for any condition. Physiological changes associated with CRMRF were more pronounced when compared to PSWT, placebo or control. Any potential stronger therapeutic benefits of CRMRF need to be confirmed by comparative clinical studies.en
dc.format.extent12
dc.format.extent985663
dc.language.isoeng
dc.relation.ispartofElectromagnetic Biology and Medicine
dc.subjectElectrophysical agents
dc.subjectphysiological effects
dc.subjectradiofrequency
dc.subjectskin blood flow
dc.subjectskin temperature
dc.subjectBiophysics
dc.subjectMedicine (miscellaneous)
dc.titleSkin thermophysiological effects of 448 kHz capacitive resistive monopolar radiofrequency in healthy adults : A randomised crossover study and comparison with Pulsed Shortwave Therapyen
dc.contributor.institutionPhysiotherapy
dc.contributor.institutionSchool of Health and Social Work
dc.contributor.institutionDepartment of Allied Health Professions, Midwifery and Social Work
dc.contributor.institutionCentre for Applied Clinical, Health and Care Research (CACHE)
dc.description.statusPeer reviewed
dc.date.embargoedUntil2019-01-08
dc.identifier.urlhttp://www.scopus.com/inward/record.url?scp=85041006256&partnerID=8YFLogxK
rioxxterms.versionofrecord10.1080/15368378.2017.1422260
rioxxterms.typeJournal Article/Review
herts.preservation.rarelyaccessedtrue


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record